PT-2015-1287 · Hospira · Hospira Lifecare Pca Infusion System

Billy Rios

·

Publicado

2015-04-29

·

Atualizado

2017-01-03

·

CVE-2015-3459

CVSS v2.0

10

Alta

VetorAV:N/AC:L/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions Hospira LifeCare PCA Infusion System versions prior to 7.0
Description The issue concerns the lack of authentication for Telnet sessions on the communication module of the affected device. This allows a remote attacker to gain control over the device by exploiting the absence of authentication on port 23, specifically for root Telnet sessions. The attacker can modify the pump configuration using unspecified commands.
Recommendations For versions prior to 7.0, consider disabling the Telnet protocol or restricting access to port 23 until a patch is available. As a temporary workaround, limit network access to the device to minimize the risk of exploitation.

Correção

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

BDU:2015-10323
CVE-2015-3459

Produtos afetados

Hospira Lifecare Pca Infusion System