PT-2015-5962 · Medicomp · Medicomp Medcin Engine

Ryan Wincey

·

Publicado

2015-10-29

·

Atualizado

2015-10-29

·

CVE-2015-2901

CVSS v2.0

6.8

Média

VetorAV:N/AC:M/Au:N/C:P/I:P/A:P
Name of the Vulnerable Software and Affected Versions Medicomp MEDCIN Engine version 2.22.20142.166
Description The issue is related to multiple stack-based buffer overflows that could allow remote attackers to execute arbitrary code. This is possible via a crafted packet on port 8190, specifically through the GetProperty info getproperty function and the GetProperty UdfCodeList function.
Recommendations For Medicomp MEDCIN Engine version 2.22.20142.166, consider restricting access to port 8190 to minimize the risk of exploitation. As a temporary workaround, disabling the info getproperty and UdfCodeList functions within the GetProperty functionality might help until a patch is available. At the moment, there is no information about a newer version that contains a fix for this vulnerability.

Buffer Overflow

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2015-2901

Produtos afetados

Medicomp Medcin Engine