PT-2015-6357 · Hospira · Hospira Lifecare Pca Infusion System

Publicado

2015-07-06

·

Atualizado

2016-12-06

·

CVE-2015-3958

CVSS v2.0

7.8

Alta

VetorAV:N/AC:L/Au:N/C:N/I:N/A:C
Name of the Vulnerable Software and Affected Versions Hospira LifeCare PCA Infusion System versions 5.0 and earlier
Description The issue allows remote attackers to cause a denial of service, resulting in a forced manual reboot, via a flood of TCP packets.
Recommendations For versions 5.0 and earlier, consider implementing network traffic filtering to restrict excessive TCP packets as a temporary mitigation measure. At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2015-3958

Produtos afetados

Hospira Lifecare Pca Infusion System