PT-2017-9633 · Cadd Solis · Cadd-Solis Medication Safety

Andrew Gothard

·

Publicado

2017-02-13

·

Atualizado

2017-02-28

·

CVE-2016-8355

CVSS v3.1

9.9

Crítica

VetorAV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions CADD-Solis Medication Safety Software versions 1.0 through 3.1
Description The issue allows an authenticated user to gain elevated privileges on the SQL database. This could enable the user to modify drug libraries, add and delete users, and change user permissions. It is noted that physical access to the pump is required to install drug library updates.
Recommendations For versions 1.0 through 3.1, restrict access to the SQL database to prevent unauthorized modifications and changes to user permissions. As a temporary workaround, consider limiting user permissions to minimize the risk of exploitation until a patch is available. At the moment, there is no information about a newer version that contains a fix for this vulnerability.

Missing Authentication

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2016-8355

Produtos afetados

Cadd-Solis Medication Safety