PT-2018-5443 · Abbott Laboratories · Pacemaker

Publicado

2018-04-25

·

Atualizado

2019-10-09

·

CVE-2017-12712

CVSS v3.1

8.8

Alta

VetorAV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions: Abbott Laboratories pacemakers versions prior to Aug 28, 2017
Description: The authentication algorithm in the pacemakers can be compromised or bypassed, allowing a nearby attacker to issue unauthorized commands via RF communications. This involves an authentication key and time stamp.
Recommendations: For pacemakers manufactured prior to Aug 28, 2017, apply the firmware update developed by Abbott to help mitigate the identified issue.

Correção

Improper Authentication

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2017-12712

Produtos afetados

Pacemaker