PT-2018-8519 · Philips · Philips Intellivue Mx40
Publicado
2018-04-30
·
Atualizado
2019-10-09
·
CVE-2017-9658
CVSS v3.1
6.5
Média
| Vetor | AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
Name of the Vulnerable Software and Affected Versions
Philips IntelliVue MX40 Version B.06.18 and earlier
Description
The issue arises when certain 802.11 network management messages unnecessarily trigger wireless access point blacklisting security defenses. This can require hospital staff to intervene, resetting the device to reestablish a network connection. During this time, the device can either connect to an alternative access point or remain in local monitoring mode until reset. The estimated number of potentially affected devices is not specified.
Recommendations
For Philips IntelliVue MX40 Version B.06.18 and earlier, apply the software update to Version B.06.18 or later to fix the issue and implement mitigations that reduce the risk associated with improper handling of exceptional conditions.
Correção
Improper Handling of Exceptional Conditions
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾
Enumeração de Fraquezas
Identificadores relacionados
Produtos afetados
Philips Intellivue Mx40