PT-2018-9987 · Medtronic · Medtronic 2090 Carelink Programmer

Publicado

2018-07-02

·

Atualizado

2019-10-09

·

CVE-2018-10596

CVSS v3.1

7.1

Alta

VetorAV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions: Medtronic 2090 CareLink Programmer all versions
Description: The issue concerns the product's update process, which uses a virtual private network connection. However, it fails to verify if it is still connected to this network before downloading updates. This could allow an attacker with local network access to influence these communications.
Recommendations: For Medtronic 2090 CareLink Programmer all versions, ensure that the device verifies its connection to the virtual private network before downloading updates to prevent potential interference from attackers. At the moment, there is no information about a newer version that contains a fix for this vulnerability.

Information Disclosure

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2018-10596

Produtos afetados

Medtronic 2090 Carelink Programmer