PT-2018-9990 · Philips · Intellivue Patient Monitors Mx+3

Publicado

2018-06-05

·

Atualizado

2021-05-10

·

CVE-2018-10599

CVSS v2.0

2.9

Baixa

VetorAV:A/AC:M/Au:N/C:P/I:N/A:N
Name of the Vulnerable Software and Affected Versions: IntelliVue Patient Monitors MP Series versions Rev B-M IntelliVue Patient Monitors MX versions Rev J-M Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0, J.3 IntelliVue Patient Monitors X3/MX100 version Rev M
Description: The issue allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.
Recommendations: For IntelliVue Patient Monitors MP Series versions Rev B-M, update to a version that includes the necessary security patches. For IntelliVue Patient Monitors MX versions Rev J-M, apply the recommended configuration changes to restrict access to the vulnerable component. For Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0, J.3, restrict network access to the device to minimize the risk of exploitation. For IntelliVue Patient Monitors X3/MX100 version Rev M, consider implementing additional security measures to prevent unauthorized access to the device.

Correção

Information Disclosure

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2018-10599

Produtos afetados

Avalon Fetal/Maternal Monitors
Intellivue Patient Monitors Mp Series
Intellivue Patient Monitors Mx
Intellivue Patient Monitors X3/Mx100