PT-2019-13406 · Medtronic · Medtronic Valleylab Exchange Client+2
Publicado
2019-11-08
·
Atualizado
2019-11-13
·
CVE-2019-13543
CVSS v3.1
7.5
Alta
| Vetor | AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
Name of the Vulnerable Software and Affected Versions
Medtronic Valleylab Exchange Client version 3.4 and below
Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below
Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below
Description
The issue concerns the use of multiple sets of hard-coded credentials in the affected software. If these credentials are discovered, they can be used to read files on the device.
Recommendations
For Medtronic Valleylab Exchange Client version 3.4 and below, update to a version above 3.4 to resolve the issue.
For Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, update to a version above 4.0.0 to resolve the issue.
For Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below, update to a version above 1.1.0 to resolve the issue.
Correção
Using Hardcoded Credentials
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾
Enumeração de Fraquezas
Identificadores relacionados
Produtos afetados
Medtronic Valleylab Exchange Client
Valleylab Ft10 Energy Platform
Valleylab Fx8 Energy Platform