PT-2019-7130 · Baxter · Baxter Sigma Spectrum Infusion System
Publicado
2019-03-26
·
Atualizado
2019-10-09
·
CVE-2014-5434
CVSS v2.0
5.0
Média
| Vetor | AV:N/AC:L/Au:N/C:P/I:N/A:N |
Name of the Vulnerable Software and Affected Versions
Baxter SIGMA Spectrum Infusion System version 6.05
Description
The issue concerns a default account with hard-coded credentials used in conjunction with the FTP protocol. It is specified that no files can be transferred to or from the wireless battery module using this account.
Recommendations
For Baxter SIGMA Spectrum Infusion System version 6.05, update to version 8, which incorporates necessary hardware and software changes to address the issue.
Correção
Using Hardcoded Credentials
Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾
Identificadores relacionados
Produtos afetados
Baxter Sigma Spectrum Infusion System