PT-2025-30629 · Medtronic · Mycarelink Patient Monitor

CVE-2025-4394

·

Publicado

2025-07-24

·

Atualizado

2025-08-25

CVSS v3.1

6.8

Média

VetorAV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions Medtronic MyCareLink Patient Monitor versions 24950 and 24952 before June 25, 2025
Description The Medtronic MyCareLink Patient Monitor utilizes an unencrypted filesystem on its internal storage. This allows an attacker with physical access to read and modify files stored on the device.
Recommendations Ensure that physical access to the Medtronic MyCareLink Patient Monitor models 24950 and 24952 is restricted before June 25, 2025.

Correção

LPE

Cleartext Storage of Sensitive Information

Encontrou algum problema na descrição? Tem algo a acrescentar? Fique à vontade para nos escrever 👾

Enumeração de Fraquezas

Identificadores relacionados

CVE-2025-4394

Produtos afetados

Mycarelink Patient Monitor