PT-2014-2064 · Zoll · Zoll Defibrillator / Monitor E Series

Published

2014-08-12

·

Updated

2017-09-08

·

CVE-2007-6756

CVSS v2.0

4.9

Medium

VectorAV:L/AC:L/Au:N/C:N/I:N/A:C
Name of the Vulnerable Software and Affected Versions ZOLL Defibrillator / Monitor M Series, E Series, and R Series (affected versions not specified)
Description The issue concerns a default password for System Configuration mode, allowing physically proximate attackers to modify device configuration. This could lead to a denial of service, potentially resulting in adverse human health effects.
Recommendations For ZOLL Defibrillator / Monitor M Series, E Series, and R Series, change the default password for System Configuration mode to prevent unauthorized access and modifications. As a temporary workaround, consider restricting physical access to the devices until a secure configuration is implemented. At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2007-6756

Affected Products

Zoll Defibrillator / Monitor E Series