PT-2015-1621 · Hospira · Hospira Lifecare Pca Infusion System

Published

2015-07-06

·

Updated

2015-07-08

·

CVE-2015-1011

CVSS v2.0

5.0

Medium

VectorAV:N/AC:L/Au:N/C:P/I:N/A:N
Name of the Vulnerable Software and Affected Versions Hospira LifeCare PCA Infusion System versions prior to 7.0
Description The issue is related to the lack of protection for service data in the device's software, which can be exploited by a remote attacker to gain access to the device. Hardcoded credentials in the system make it easier for attackers to obtain access.
Recommendations For versions prior to 7.0, update to version 7.0 or later to resolve the issue. As a temporary workaround, consider restricting access to the device to minimize the risk of exploitation.

Fix

Information Disclosure

Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

BDU:2015-10956
CVE-2015-1011

Affected Products

Hospira Lifecare Pca Infusion System