PT-2015-3917 · Ge Healthcare · Ge Healthcare Precision Thunis-800+

Published

2015-08-04

·

Updated

2018-03-28

·

CVE-2014-7233

CVSS v2.0

10

High

VectorAV:N/AC:L/Au:N/C:C/I:C/A:C
Name of the Vulnerable Software and Affected Versions GE Healthcare Precision THUNIS-800+ (affected versions not specified)
Description The issue concerns default passwords used in various installation and setup processes. Specifically, the default passwords are: (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740 122 Setup.exe, (3) hrml for "Setup and Activation" using DSASetup, and (4) an empty string for Shutter Configuration. The impact and attack vectors of these default passwords are unspecified.
Recommendations At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2014-7233

Affected Products

Ge Healthcare Precision Thunis-800+