PT-2015-5962 · Medicomp · Medicomp Medcin Engine
Ryan Wincey
·
Published
2015-10-29
·
Updated
2015-10-29
·
CVE-2015-2901
CVSS v2.0
6.8
Medium
| Vector | AV:N/AC:M/Au:N/C:P/I:P/A:P |
Name of the Vulnerable Software and Affected Versions
Medicomp MEDCIN Engine version 2.22.20142.166
Description
The issue is related to multiple stack-based buffer overflows that could allow remote attackers to execute arbitrary code. This is possible via a crafted packet on port 8190, specifically through the
GetProperty info getproperty function and the GetProperty UdfCodeList function.Recommendations
For Medicomp MEDCIN Engine version 2.22.20142.166, consider restricting access to port 8190 to minimize the risk of exploitation. As a temporary workaround, disabling the
info getproperty and UdfCodeList functions within the GetProperty functionality might help until a patch is available. At the moment, there is no information about a newer version that contains a fix for this vulnerability.Buffer Overflow
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Medicomp Medcin Engine