Name of the Vulnerable Software and Affected Versions Hospira LifeCare PCA Infusion System versions 5.0 and earlier
Description The issue allows remote attackers to cause a denial of service, resulting in a forced manual reboot, via a flood of TCP packets.
Recommendations For versions 5.0 and earlier, consider implementing network traffic filtering to restrict excessive TCP packets as a temporary mitigation measure. At the moment, there is no information about a newer version that contains a fix for this vulnerability.