PT-2018-16962 · Medtronic · Medtronic 2090 Carelink Programmer

Billy Rios

+1

·

Published

2018-05-04

·

Updated

2019-10-09

·

CVE-2018-5446

CVSS v3.1

4.9

Medium

VectorAV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N
Name of the Vulnerable Software and Affected Versions Medtronic 2090 Carelink Programmer (affected versions not specified)
Description The issue concerns a per-product username and password stored in a recoverable format. This could allow an attacker with physical access to the programmer to obtain credentials to the software deployment network.
Recommendations At the moment, there is no information about a newer version that contains a fix for this vulnerability.

Insufficiently Protected Credentials

Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2018-5446

Affected Products

Medtronic 2090 Carelink Programmer