PT-2018-18680 · Medtronic · Medtronic Mycarelink Patient Monitor+2
Peter Morgan
·
Published
2018-07-02
·
Updated
2019-10-09
·
CVE-2018-8870
CVSS v3.1
6.4
Medium
| Vector | AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
Name of the Vulnerable Software and Affected Versions
Medtronic MyCareLink Patient Monitor versions all
Medtronic 24950 MyCareLink Monitor versions all
Medtronic 24952 MyCareLink Monitor versions all
Description
The issue concerns a hard-coded operating system password in the affected devices. An attacker who gains physical access to the device can exploit this by removing the case, connecting to the debug port, and using the password to gain privileged access to the operating system.
Recommendations
For Medtronic MyCareLink Patient Monitor, consider restricting physical access to the device to minimize the risk of exploitation.
For Medtronic 24950 MyCareLink Monitor, restrict physical access to the device until a fix is available.
For Medtronic 24952 MyCareLink Monitor, limit access to the debug port to prevent unauthorized use of the hard-coded password.
Fix
Using Hardcoded Credentials
Found an issue in the description? Have something to add? Feel free to write us 👾
Related Identifiers
Affected Products
Medtronic 24950 Mycarelink Monitor
Medtronic 24952 Mycarelink Monitor
Medtronic Mycarelink Patient Monitor