PT-2018-5450 · Smiths Medical · Medfusion 4000 Wireless Syringe Infusion Pump
Published
2018-02-15
·
Updated
2018-03-02
·
CVE-2017-12722
CVSS v3.1
5.3
Medium
| Vector | AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L |
Name of the Vulnerable Software and Affected Versions:
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump versions 1.1, 1.5, 1.6
Description:
An Out-of-bounds Read issue was discovered in the pump, where a third-party component reads memory out of bounds, causing the communications module to crash. The crash of the communications module would not impact the operation of the therapeutic module.
Recommendations:
For version 1.1, update the third-party component to prevent the out-of-bounds read issue.
For version 1.5, update the third-party component to prevent the out-of-bounds read issue.
For version 1.6, update the third-party component to prevent the out-of-bounds read issue.
Fix
Out of bounds Read
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Medfusion 4000 Wireless Syringe Infusion Pump