PT-2018-5450 · Smiths Medical · Medfusion 4000 Wireless Syringe Infusion Pump

Published

2018-02-15

·

Updated

2018-03-02

·

CVE-2017-12722

CVSS v3.1

5.3

Medium

VectorAV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L
Name of the Vulnerable Software and Affected Versions: Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump versions 1.1, 1.5, 1.6
Description: An Out-of-bounds Read issue was discovered in the pump, where a third-party component reads memory out of bounds, causing the communications module to crash. The crash of the communications module would not impact the operation of the therapeutic module.
Recommendations: For version 1.1, update the third-party component to prevent the out-of-bounds read issue. For version 1.5, update the third-party component to prevent the out-of-bounds read issue. For version 1.6, update the third-party component to prevent the out-of-bounds read issue.

Fix

Out of bounds Read