PT-2019-13406 · Medtronic · Medtronic Valleylab Exchange Client+2

Published

2019-11-08

·

Updated

2019-11-13

·

CVE-2019-13543

CVSS v3.1

7.5

High

VectorAV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Name of the Vulnerable Software and Affected Versions Medtronic Valleylab Exchange Client version 3.4 and below Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below
Description The issue concerns the use of multiple sets of hard-coded credentials in the affected software. If these credentials are discovered, they can be used to read files on the device.
Recommendations For Medtronic Valleylab Exchange Client version 3.4 and below, update to a version above 3.4 to resolve the issue. For Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, update to a version above 4.0.0 to resolve the issue. For Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below, update to a version above 1.1.0 to resolve the issue.

Fix

Using Hardcoded Credentials

Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2019-13543

Affected Products

Medtronic Valleylab Exchange Client
Valleylab Ft10 Energy Platform
Valleylab Fx8 Energy Platform