PT-2020-14450 · Nescomed · Nescomed Multipara Monitor M1000
Arun Magesh
·
Published
2020-08-26
·
Updated
2024-02-14
·
CVE-2020-15483
CVSS v2.0
7.2
High
| Vector | AV:L/AC:L/Au:N/C:C/I:C/A:C |
Name of the Vulnerable Software and Affected Versions
Nescomed Multipara Monitor M1000 devices (affected versions not specified)
Description
An issue was discovered where the physical UART debug port on the devices provides a shell without requiring a password, giving complete access.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Exploit
Missing Authentication
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Nescomed Multipara Monitor M1000