PT-2021-20349 · B. Braun · B. Braun Spacecom2

Published

2021-08-25

·

Updated

2021-09-01

·

CVE-2021-33882

CVSS v3.1

8.6

High

VectorAV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N
Name of the Vulnerable Software and Affected Versions: B. Braun SpaceCom2 versions prior to 012U000062
Description: A Missing Authentication for Critical Function issue allows a remote attacker to reconfigure the device from an unknown source because of a lack of authentication on proprietary networking commands.
Recommendations: For versions prior to 012U000062, update to version 012U000062 or later to resolve the issue. As a temporary workaround, consider restricting access to proprietary networking commands until a patch is applied.

Fix

Missing Authentication

Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2021-33882

Affected Products

B. Braun Spacecom2