PT-2021-21541 · Swisslog · Swisslog Healthcare Nexus Panel+1
Published
2021-08-02
·
Updated
2021-08-10
·
CVE-2021-37161
CVSS v3.1
9.8
Critical
| Vector | AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
Name of the Vulnerable Software and Affected Versions
Swisslog Healthcare Nexus Panel versions prior to 7.2.5.7
Description
A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel. This issue allows an attacker to overwrite an internal queue data structure, which can lead to remote code execution.
Recommendations
For versions prior to 7.2.5.7, update to version 7.2.5.7 or later to resolve the issue. As a temporary workaround, consider restricting access to the HMI3 Control Panel to minimize the risk of exploitation.
Fix
Buffer Overflow
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Hmi3 Control Panel