PT-2021-21542 · Swisslog · Swisslog Healthcare Nexus Panel
Published
2021-08-02
·
Updated
2021-08-10
·
CVE-2021-37162
CVSS v3.1
9.8
Critical
| Vector | AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
Name of the Vulnerable Software and Affected Versions
Swisslog Healthcare Nexus Panel versions prior to 7.2.5.7
Description
A buffer overflow issue was discovered in the HMI3 Control Panel. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.
Recommendations
For versions prior to 7.2.5.7, update to Nexus Software 7.2.5.7 or later to resolve the issue. As a temporary workaround, consider restricting access to the UDP endpoint to minimize the risk of exploitation.
Fix
Buffer Overflow
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Swisslog Healthcare Nexus Panel