PT-2021-22112 · Boston Scientific · Boston Scientific Zoom Latitude Model 3120
Christian Dresen
+3
·
Published
2021-10-04
·
Updated
2021-10-13
·
CVE-2021-38400
CVSS v3.1
6.9
Medium
| Vector | AV:P/AC:H/PR:N/UI:R/S:C/C:H/I:H/A:L |
Name of the Vulnerable Software and Affected Versions
Boston Scientific Zoom Latitude Model 3120
Description
An attacker with physical access to the device can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
Recommendations
For Boston Scientific Zoom Latitude Model 3120, consider restricting physical access to the device to minimize the risk of exploitation. As a temporary workaround, limit the use of removable storage devices and ensure that all USB ports are securely managed to prevent the creation of specially crafted USB devices. At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Boston Scientific Zoom Latitude Model 3120