PT-2022-8704 · B. Braun Melsungen Ag · Data Module Compactplus+1

Birk Kauer

+3

·

Published

2022-04-14

·

Updated

2022-04-21

·

CVE-2020-25162

CVSS v2.0

7.8

High

VectorAV:N/AC:L/Au:N/C:C/I:N/A:N
Name of the Vulnerable Software and Affected Versions B. Braun Melsungen AG SpaceCom versions L81/U61 and earlier B. Braun Melsungen AG Data module compactplus versions A10 and A11
Description A XPath injection issue allows unauthenticated remote attackers to access sensitive information and escalate privileges.
Recommendations For B. Braun Melsungen AG SpaceCom versions L81/U61 and earlier, update to a version later than L81/U61 to resolve the issue. For B. Braun Melsungen AG Data module compactplus versions A10 and A11, update to a version later than A11 to resolve the issue. As a temporary workaround, consider restricting access to sensitive information and privileges until a patch is available.

Fix

Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2020-25162

Affected Products

Data Module Compactplus
Spacecom