PT-2022-8709 · B. Braun Melsungen Ag · Data Module Compactplus+1
Published
2022-04-14
·
Updated
2022-04-21
·
CVE-2020-25168
CVSS v3.1
3.3
Low
| Vector | AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N |
Name of the Vulnerable Software and Affected Versions
B. Braun Melsungen AG SpaceCom versions L81/U61 and earlier
B. Braun Melsungen AG Data module compactplus versions A10 and A11
Description
The issue concerns hard-coded credentials that allow attackers with command line access to access the device's Wi-Fi module.
Recommendations
For B. Braun Melsungen AG SpaceCom versions L81/U61 and earlier, consider changing the hard-coded credentials to unique, secure credentials.
For B. Braun Melsungen AG Data module compactplus versions A10 and A11, consider changing the hard-coded credentials to unique, secure credentials.
As a temporary workaround, consider restricting command line access to the device until a patch is available.
Fix
Using Hardcoded Credentials
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Data Module Compactplus
Spacecom