PT-2024-33302 · Unknown · Ventilator Software Tools
Published
2024-11-14
·
Updated
2024-11-15
·
CVE-2024-48966
CVSS v3.1
10
Critical
| Vector | AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
Name of the Vulnerable Software and Affected Versions
Ventilator software tools (affected versions not specified)
Description
The software tools used by service personnel to test and calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Missing Authentication
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products