PT-2025-5994 · Unknown · Eli 250C/Bur 250C Resting Electrocardiograph+3

Published

2025-02-07

·

Updated

2025-02-07

·

CVE-2022-26389

CVSS v3.1

7.7

High

VectorAV:N/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:H
Name of the Vulnerable Software and Affected Versions ELI 380 Resting Electrocardiograph versions 2.6.0 and prior ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph versions 2.3.1 and prior ELI 250c/BUR 250c Resting Electrocardiograph versions 2.1.2 and prior ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph versions 2.2.0 and prior
Description An improper access control issue may allow privilege escalation.
Recommendations For ELI 380 Resting Electrocardiograph versions 2.6.0 and prior, update to a version later than 2.6.0 to resolve the issue. For ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph versions 2.3.1 and prior, update to a version later than 2.3.1 to resolve the issue. For ELI 250c/BUR 250c Resting Electrocardiograph versions 2.1.2 and prior, update to a version later than 2.1.2 to resolve the issue. For ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph versions 2.2.0 and prior, update to a version later than 2.2.0 to resolve the issue.

Fix

LPE

Improper Access Control

Weakness Enumeration

Related Identifiers

CVE-2022-26389

Affected Products

Eli 150C/Bur 150C/Mlbur 150C Resting Electrocardiograph
Eli 250C/Bur 250C Resting Electrocardiograph
Eli 280/Bur280/Mlbur 280 Resting Electrocardiograph
Eli 380 Resting Electrocardiograph