PT-2026-43357 · Eppendorf · Bioflo 320

CVE-2026-7251

·

Published

2026-05-26

·

Updated

2026-07-23

CVSS v3.1

9.8

Critical

VectorAV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Name of the Vulnerable Software and Affected Versions Eppendorf BioFlo 320 (affected versions not specified)
Description The VNC server uses a hard-coded password. A remote attacker who knows the network address of a device with remote access enabled can use this password to gain full control of the user interface and all control panel features. VNC traffic is not encrypted.
Recommendations At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2026-7251

Affected Products

Bioflo 320