PT-2026-45564 · Dräger · Infinity Kappa+2
CVE-2019-25716
·
Published
2026-06-01
·
Updated
2026-07-22
CVSS v3.1
7.5
High
| Vector | AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
Name of the Vulnerable Software and Affected Versions
Dräger Infinity Delta (affected versions not specified)
Dräger Infinity Delta XL (affected versions not specified)
Dräger Infinity Kappa (affected versions not specified)
Description
A denial-of-service issue exists where remote attackers can cause the monitor to reboot by sending a malformed network packet. Repeated transmission of these packets can disrupt patient monitoring, eventually forcing the device to revert to its default configuration and lose network connectivity.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
DoS
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Infinity Delta
Infinity Delta Xl
Infinity Kappa