PT-2026-45627 · Dräger · Infinity Explorer C700

Published

2026-06-01

·

Updated

2026-06-01

·

CVE-2019-25718

CVSS v3.1

8.4

High

VectorAV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.

Fix

UI Misrepresentation of Critical Information

Weakness Enumeration

Related Identifiers

CVE-2019-25718

Affected Products