PT-2026-45739 · Dräger · Infinity Acute Care System+1
CVE-2019-25719
·
Published
2026-06-02
·
Updated
2026-06-02
CVSS v3.1
8.6
High
| Vector | AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
Name of the Vulnerable Software and Affected Versions
Dräger Infinity Acute Care System versions prior to VG4.1.2
Standalone Infinity M540 versions prior to VG4.1.2
Description
Network message handling issues allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings, including alarm states or alarm limits, and overwhelm the system with incoming data, leading to device reboots and loss of network functionality.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Infinity Acute Care System
Standalone Infinity M540 Patient Monitor