PT-2026-45739 · Dräger · Infinity Acute Care System+1

Published

2026-06-02

·

Updated

2026-06-02

·

CVE-2019-25719

CVSS v3.1

8.6

High

VectorAV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Fix

Weakness Enumeration

Related Identifiers

CVE-2019-25719

Affected Products

Infinity Acute Care System
Standalone Infinity M540 Patient Monitor