PT-2026-45739 · Dräger · Infinity Acute Care System+1
Published
2026-06-02
·
Updated
2026-06-02
·
CVE-2019-25719
CVSS v3.1
8.6
High
| Vector | AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Fix
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Infinity Acute Care System
Standalone Infinity M540 Patient Monitor