PT-2026-45810 · Dräger · Infinity M300

Published

2026-06-02

·

Updated

2026-06-02

·

CVE-2019-25721

CVSS v3.1

6.5

Medium

VectorAV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.

Fix

Resource Exhaustion

Weakness Enumeration

Related Identifiers

CVE-2019-25721

Affected Products

Infinity M300