PT-2026-45811 · Dräger · Sc 6002Xl+4
Published
2026-06-02
·
Updated
2026-06-02
·
CVE-2019-25722
CVSS v3.1
7.6
High
| Vector | AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Fix
Using Hardcoded Credentials
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Sc 6002Xl
Sc 7000
Sc6802Xl
Sc8000
Sc90000 Xl