PT-2026-45812 · Dräger · Perseus A500
CVE-2019-25723
·
Published
2026-06-02
·
Updated
2026-06-04
CVSS v3.1
4.0
Medium
| Vector | AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L |
Name of the Vulnerable Software and Affected Versions
Dräger Perseus A500 versions 2.00 through 2.02
Description
Improper input handling allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. This can overload the internal processor with malformed data, triggering a warm restart. Consequently, ventilation pressure drops to ambient level, interrupting ventilation for several seconds before therapy resumes.
Recommendations
At the moment, there is no information about a newer version that contains a fix for this vulnerability.
DoS
Found an issue in the description? Have something to add? Feel free to write us 👾
Weakness Enumeration
Related Identifiers
Affected Products
Perseus A500