PT-2026-45813 · Dräger · Infinity M300

Published

2026-06-02

·

Updated

2026-06-02

·

CVE-2019-25724

CVSS v3.1

6.5

Medium

VectorAV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Fix

Resource Exhaustion

Weakness Enumeration

Related Identifiers

CVE-2019-25724

Affected Products

Infinity M300