PT-2026-45815 · Dräger · Atlan A350

Published

2026-06-02

·

Updated

2026-06-04

·

CVE-2021-4479

CVSS v3.1

4.0

Medium

VectorAV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Name of the Vulnerable Software and Affected Versions Dräger Atlan A350 versions 1.00 through 1.01
Description Improper input handling allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Transmitting malformed data can overload the internal processor, which gradually disrupts device operation over several hours. This results in loss of data transmission, delayed display of real-time curves, and a deviation between displayed airway pressure values and screen curves.
Recommendations At the moment, there is no information about a newer version that contains a fix for this vulnerability.

DoS

Found an issue in the description? Have something to add? Feel free to write us 👾

Weakness Enumeration

Related Identifiers

CVE-2021-4479

Affected Products

Atlan A350